SHENZHEN DAFORD ELECTRIC APPLIANCE CO., LTD

12W IEC 60601 medical standard 3.7V2A 5V2A Interchangeable medical power supply with CE TUV MARK NRTL

DP-MI12 Brand:DAFORD OR CUSTOMER SUPPLY IN PUT: 90-264VAC OUT PUT: 5-24V/0.1-2A CERTIFICATE: TUV MARK .UL. NRTL. CE

SEND INQUIRY

You can make an inquiry about this product.
Your e-mail will not be leaked.

  • PRODUCT DETAIL

IEC 60601 Precision medical equipment power supply

3.7V2A  medical power supply

5V2A medical power supply

This series is wall mount and interchangeable type adaptor for use in medical equipment. This series is EN/IEC60601-1 and EN/IEC60601-1-11 approved. The specialty of this power supply is the 2MOPP (two means of patient protection) safety approval and the fulfillment of the required EMC tests for home and professional healthcare environments.

Item DP-MI unit notes
005100 005200 006100 006200 009133
Rated output voltage +5 +5 +6 +6 +9 V  
Output voltage range 4.70~5.40 4.70~5.40 5.7~6.4 5.7~6.4 8.7~9.4 V  
Rated output current 0~1 0~2 0~1 0~2 0~1.33 A  
Rated output power 5 10 6 12 12 W  
load regulation ≤±5 ≤±5 ≤±5 ≤±5 ≤±5 %
voltage regulation ≤±0.5 ≤±0.5 ≤±0.5 ≤±0.5 ≤±0.5 %  
Rise time ≤30 ≤30 ≤30 ≤30 ≤30 mS  
overshoot ≤10 ≤10 ≤10 ≤10 ≤10 %  
Hold up time ≥5 ≥5 ≥5 ≥5 ≥5 mS @220VAC
Ripple and noise The ripples and noises are less than 80mV for 5V @6V and 9V series power outputs, and less than 120mV for 12V and 15V series power supply outputs which measured by 20MHz bandwidth oscilloscope and terminated each output with a 0.1uF ceramic capacitor and a 10uF aluminum electrolytic capacitor .
Over voltage protection The power supply shall be hiccupped when output voltage reaches its over-voltage condition trigger point, and when the fault condition is removed, the power supply shall not be self-recovering. The over voltage protection function is built-in the PWM IC of the power supply.
Over current protection ≤2 ≤3 ≤2 ≤3 ≤2 A  
Over temperature protection The power supply shall go into latch-off mode when the IC’s temperature is over its trigger point ,and when IC temperature is lower than regulation value, the power supply shall be self-recovering.
Short circuit protection The power supply shall withstand continuous current of short circuit on output any damage, and when the fault condition is removed, the power supply shall be self-recovering.

 

Item DP-MI unit notes
012100 015080      
Rated output voltage +12 +15       V  
Output voltage range 11.7~12.4 14.6~15.4       V  
Rated output current 0~1 0~0.8       A  
Rated output power 12 12 W  
load regulation ≤±5 ≤±5       %  
voltage regulation ≤±0.5 ≤±0.5       %  
Rise time ≤30 ≤30       mS  
overshoot ≤10 ≤10       %  
Hold up time ≥5 ≥5 mS @220VAC
Ripple and noise The ripples and noises are less than 80mV for 5V @6V and 9V series power outputs, and less than 120mV for 12V and 15V series power supply outputs which measured by 20MHz bandwidth oscilloscope and terminated each output with a 0.1uF ceramic capacitor and a 10uF aluminum electrolytic capacitor .
Over voltage protection The power supply shall be hiccupped when output voltage reaches its over-voltage condition trigger point, and when the fault condition is removed, the power supply shall not be self-recovering. The over voltage protection function is built-in the PWM IC of the power supply.
Over current protection ≤2 ≤1.5       A  
Over temperature protection The power supply shall go into latch-off mode when the IC’s temperature is over its trigger point ,and when IC temperature is lower than regulation value, the power supply shall be self-recovering.
Short circuit protection The power supply shall withstand continuous current of short circuit on output any damage, and when the fault condition is removed, the power supply shall be self-recovering.

 